Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C8RMS12 - alkaline phosphatase (ALP) in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC8RMS12
Reference Laboratory StatusMedical School Hannover
Analyte Namealkaline phosphatase (ALP)
IUPAC IFCC NumberEC 3.1.3.1
Applicable Matricesmatrix of calibrators and control materials
Traceable to SIyes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureSpectrophotometry for alkaline phosphatase (ALP) in blood serum
JCTLM Database Identifier NumberC7RMP2R
Procedure Difference in Your Laboratoryno
QuantityCatalytic activity concentration
From (0)0.092
To (0)11.3
Unit (0)µkat/l
From (1)2.8
To (1)2.8
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from 0.092 µkal/l (5.5 U/L) to 11.3 µkal/l (680 U/L).
Comments on Uncertainty RangeThe expanded uncertainty depends on the matrix of the sample
Survey OrganisationJCTLM, IFCC, RfB
Intercomparison Results Publicly Availableyes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Remarks11.0
Organization AcronymRfB
Organization NameReference Institute for Bioanalytics, Calibration Laboratory II
Organization Websitehttps://rfb.bio/cgi/switchLang?lang=en
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameEnzymes