JCTLM Database: higher-order reference materials, methods and services
Details : C8RMS6 - lactate dehydrogenase (LDH) in High purity material, Calibration solution, Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C8RMS6 |
Reference Laboratory Status | Hospital |
Analyte Name | lactate dehydrogenase (LDH) |
IUPAC IFCC Number | EC 1.1.1.27 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Kinetic spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | IFCC reference
measurement
procedure (37?)
for LDH |
JCTLM Database Identifier Number | NRMeth 66 |
Procedure Difference in Your Laboratory | No |
Quantity | Catalytic activity concentration |
From (0) | 1.019 |
To (0) | 9.552 |
Unit (0) | µkat/l |
From (1) | 2.5 |
To (1) | 1.6 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The measurement range spans from 1.019 µkat/l to 9.552 µkat/l without dilution of the sample.
The range of concentration is adapted to the usual concentrations measured.
The conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L |
Reference Procedure Status Acceptable | YES |
Laboratory Status | (b) |
Survey Organisation | IFCC,www.dgkl-rfb.de:81 |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Additional Information 1 | IFCC:RELA2006~2010 IFCC:RELA2012 |
Participation Status in Intercomparison Surveys acceptable | YES |
Organization Acronym | HT711 |
Organization Name | Beijing Aerospace General Hospital Reference Laboratory |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Enzymes |