JCTLM Database: higher-order reference materials, methods and services
Details : C8RMS9 - glucose in Blood serum, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C8RMS9 |
| Reference Laboratory Status | Hospital |
| Analyte Name | glucose |
| IUPAC IFCC Number | CAS50-99-7 |
| Applicable Matrices | lyophilized, fresh or frozen human serum and calibration solution |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | CDC Hexokinase reference method for glucose |
| JCTLM Database Identifier Number | NRMeth 34 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Mass concentration |
| From (0) | 39.5 |
| To (0) | 373.4 |
| Unit (0) | mg/dl |
| From (1) | 3.3 |
| To (1) | 1.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Concentration Range | The measurement range spans from 39.5 mg/dl to 373.4 mg/dl without dilution of the sample.
The range of concentration is adapted to the usual concentrations measured. |
| Comments on Uncertainty Range | The expanded uncertainty values do not relate to the lower and higher limits of determination that are in accordance with the reference method. |
| Reference Procedure Status Acceptable | Yes |
| Laboratory Status | (b) Accredited (ISO 17025+15195) |
| Survey Organisation | IFCC RELA |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results | Hu Bin et al., Chinese Journal of Clinical Laboratory Science, 2013, 31(1), 68-73, The effect of the portable blood glucose meter test strips on blood glucose measurement results
|
| Additional Information 1 | IFCC:RELA2007~2012 |
| Participation Status in Intercomparison Surveys acceptable | Yes |
| Comments of Review Teams | Recommended for Listing ! |
| Organization Acronym | HT711 |
| Organization Name | Beijing Aerospace General Hospital Reference Laboratory |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Metabolites and substrates |