Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C9RMP1 - ID/SERS reference measurement method for serum creatinine

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC9RMP1
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameID/SERS reference measurement method for serum creatinine
Organization that developed Validated MethodPhysikalisch-Technische Bundesanstalt (PTB)
Measurement Techniques UsedIsotope dilution surface enhanced Raman scattering (ID/SERS)
Comments 2Calibrate with SRM 914a, high purity creatinine
QuantityAmount-of-substance concentration
From (0)10
To (0)100
Unit (0)µg/ml
From (1)2
To (1)5
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationStosch et al., Anal. Chem., 2005, 77, 7386-7392
Zakel et. al., Analyst, 2011, 136, 3956-3961
Analyte Category NameMetabolites and substrates
Non-CCQM Key Comparison ReportIFCC-RELA 2010 EURAMET TC-MC Project 1176
Non-CCQM Key Comparison Report Hyperlink
Measurement principle/techniqueIsotope dilution surface enhanced raman scattering