JCTLM Database: higher-order reference materials, methods and services

Details : C9RMS12 - alpha-amylase in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC9RMS12
Reference Laboratory StatusManufacturer
Analyte Namealpha-amylase
IUPAC IFCC NumberEC 3.2.1.1
Applicable Matriceslyophilized blood serum, or plasma
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC reference measurement procedure (37 °C) for amylase
JCTLM Database Identifier NumberNRMeth 70
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)1.15
To (0)12.52
Unit (0)µkat/l
From (1)2.7
To (1)2.7
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L.
The range of concentration is within the concentrations measured without dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Reference Procedure Status AcceptableThe reported (accredited) measuring range for AMY and LDH is significantly narrow, if compared with the original IFCC reference procedures. The accreditation body (here CNAS) has not certified the upper limits of the measuring ranges which are expected with measurements according to the IFCC reference measurement procedure. The issue of new certificates might be a formal action or requires additional experiments. However, a decision on this has to be made by the accreditation experts sitting in the JCTLM Executive,
Laboratory Status(b)
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYES
Comments of Review TeamsThe frequency of measurements was not reported on the excel worksheet.
Organization AcronymMaccura
Organization NameMaccura Biotechnology Co., Ltd. Reference System Dept.
Organization Websitehttps://www.maccura.com/en/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes