JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS14 - glucose in Blood serum, Blood plasma, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C9RMS14 |
Reference Laboratory Status | Manufacturer |
Analyte Name | glucose |
IUPAC IFCC Number | CAS50-99-7 |
Applicable Matrices | lyophilized blood serum, or plasma |
Traceable to SI | Yes |
Principle of Analytical Procedure | Spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC Hexokinase reference
method for glucose |
JCTLM Database Identifier Number | NRMeth 34 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 4.2 |
To (0) | 13.77 |
Unit (0) | mmol/L |
From (1) | 2.2 |
To (1) | 1.6 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | Yes(b) |
Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing . |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |