JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS14 - glucose in Blood serum, Blood plasma, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C9RMS14 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | glucose |
| IUPAC IFCC Number | CAS50-99-7 |
| Applicable Matrices | lyophilized blood serum, or plasma |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | CDC Hexokinase reference
method for glucose |
| JCTLM Database Identifier Number | NRMeth 34 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount-of-substance concentration |
| From (0) | 4.2 |
| To (0) | 13.77 |
| Unit (0) | mmol/L |
| From (1) | 2.2 |
| To (1) | 1.6 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
| Reference Procedure Status Acceptable | Yes |
| Laboratory Status | Yes(b) |
| Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Participation Status in Intercomparison Surveys acceptable | Yes |
| Comments of Review Teams | Recommended for Listing . |
| Organization Acronym | Maccura |
| Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
| Organization Website | https://www.maccura.com/en/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Metabolites and substrates |