JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS15 - urea in Blood serum, Blood plasma, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C9RMS15 |
Reference Laboratory Status | Manufacturer |
Analyte Name | urea |
IUPAC IFCC Number | CAS57-13-6 |
Applicable Matrices | lyophilized blood serum, or plasma |
Traceable to SI | Yes |
Principle of Analytical Procedure | Spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC reference method for urea |
JCTLM Database Identifier Number | NRMeth 36 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 8.29 |
To (0) | 28.67 |
Unit (0) | mmol/L |
From (1) | 2.4 |
To (1) | 1.9 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | Yes(b) |
Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing . |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |