JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS23 - uric acid in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C9RMS23 |
Reference Laboratory Status | EQAS Organizer |
Analyte Name | uric acid |
IUPAC IFCC Number | CAS69-93-2 |
Applicable Matrices | fresh, frozen or lyophilized serum |
Traceable to SI | Yes |
Principle of Analytical Procedure | GC-ID/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | University of Ghent reference method for serum uric acid |
JCTLM Database Identifier Number | NRMeth 3 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 50 |
To (0) | 1000 |
Unit (0) | µmol/L |
From (1) | 1 |
To (1) | 1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The stated expanded uncertainty value corresponds to the best measurement capability. |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | Yes(b) |
Survey Organisation | IFCC RELA Surveys |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing . |
Organization Acronym | INSTAND e. V. |
Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
Organization Website | http://www.instand-ev.de |
Organization Country ISO Code | DE |
Organization Country Name | Germany |
Analyte Category Name | Metabolites and substrates |