Details : C9RMS29 - HDL cholesterol in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC9RMS29
Reference Laboratory StatusNational Metrology Institute
Analyte NameHDL cholesterol
IUPAC IFCC NumberHDL cholesterol
Applicable Matricesfresh, or frozen serum
Traceable to SIYes
Principle of Analytical ProcedureBeta Quantification with cholesterol measured by ID-GC/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureCDC reference method for HDL cholesterol in serum
JCTLM Database Identifier NumberCYC_II_L_I_RMP_04
Procedure Difference in Your LaboratoryYES
QuantityAmount-of-substance concentration
From (0)1
To (0)10
Unit (0)mmol/L
From (1)2
To (1)5
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is relative
Reference Procedure Status AcceptableYes(method combinesCDCand DGKC methods which are listed in JCTLM database.)
Laboratory StatusYes(b)
Intercomparison Results Publicly AvailableNo
Link to Comparison ResultsThe results of the bilateral comparison between the LNE and the CDC are described in pages 104 to 107 of the PhD Thesis by Dr Maud HEUILLET
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsRecommended for Listing after the removal of lyophilized matrix
Organization AcronymLNE
Organization NameLaboratoire National de Métrologie et d'Essais
Organization Websitehttp://www.lne.fr
Organization Country ISO CodeFR
Organization Country NameFrance
Analyte Category NameMetabolites and substrates