JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS30 - LDL cholesterol in Blood serum, Blood plasma
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C9RMS30 |
| Reference Laboratory Status | National Metrology Institute |
| Analyte Name | LDL cholesterol |
| IUPAC IFCC Number | LDL cholesterol |
| Applicable Matrices | fresh, or frozen serum |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Beta Quantification with cholesterol measured by ID-GC/MS |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | CDC reference method for LDL cholesterol in serum |
| JCTLM Database Identifier Number | CYC_II_L_I_RMP_05 |
| Procedure Difference in Your Laboratory | YES |
| Quantity | Amount-of-substance concentration |
| From (0) | 1 |
| To (0) | 10 |
| Unit (0) | mmol/L |
| From (1) | 2 |
| To (1) | 5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty is relative |
| Reference Procedure Status Acceptable | Yes(method combinesCDCand DGKC methods which are listed in JCTLM database.) |
| Laboratory Status | Yes(b) |
| Intercomparison Results Publicly Available | No |
| Link to Comparison Results | The results of the bilateral comparison between the LNE and the CDC are described in pages 104 to 107 of the PhD Thesis by Dr Maud HEUILLET
|
| Participation Status in Intercomparison Surveys acceptable | Yes |
| Comments of Review Teams | Recommended for Listing after the removal of lyophilized matrix |
| Organization Acronym | LNE |
| Organization Name | Laboratoire National de Métrologie et d'Essais |
| Organization Website | http://www.lne.fr |
| Organization Country ISO Code | FR |
| Organization Country Name | France |
| Analyte Category Name | Metabolites and substrates |