JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS30 - LDL cholesterol in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C9RMS30 |
Reference Laboratory Status | National Metrology Institute |
Analyte Name | LDL cholesterol |
IUPAC IFCC Number | LDL cholesterol |
Applicable Matrices | fresh, or frozen serum |
Traceable to SI | Yes |
Principle of Analytical Procedure | Beta Quantification with cholesterol measured by ID-GC/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC reference method for LDL cholesterol in serum |
JCTLM Database Identifier Number | CYC_II_L_I_RMP_05 |
Procedure Difference in Your Laboratory | YES |
Quantity | Amount-of-substance concentration |
From (0) | 1 |
To (0) | 10 |
Unit (0) | mmol/L |
From (1) | 2 |
To (1) | 5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is relative |
Reference Procedure Status Acceptable | Yes(method combinesCDCand DGKC methods which are listed in JCTLM database.) |
Laboratory Status | Yes(b) |
Intercomparison Results Publicly Available | No |
Link to Comparison Results | The results of the bilateral comparison between the LNE and the CDC are described in pages 104 to 107 of the PhD Thesis by Dr Maud HEUILLET
|
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing after the removal of lyophilized matrix |
Organization Acronym | LNE |
Organization Name | Laboratoire National de Métrologie et d'Essais |
Organization Website | http://www.lne.fr |
Organization Country ISO Code | FR |
Organization Country Name | France |
Analyte Category Name | Metabolites and substrates |