Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C9RMS31 - HbA1c in Other, Whole blood

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC9RMS31
Reference Laboratory StatusNational Metrology Institute
Analyte NameHbA1c
IUPAC IFCC NumberQU60433
Applicable Matriceslyophilized, fresh, or frozen human hemolysate
Traceable to SIYes
Principle of Analytical ProcedureESI-LC/MS
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC method for HbA1c (HPLC / MS-ESI)
JCTLM Database Identifier NumberC1RMP_P16_MS
Procedure Difference in Your LaboratoryYES
QuantityAmount-of-substance fraction of [HbA1c/(HbA1c+HbA0)] expressed in mmol/mol
From (0)1
To (0)15
Unit (0)%
From (1)2
To (1)4
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is relative
Reference Procedure Status AcceptableYes
Laboratory StatusAccredited
Survey OrganisationIFCC RELA 2010 (Lab 105)
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsLNE's acceptable.
Organization AcronymLNE
Organization NameLaboratoire National de Métrologie et d'Essais
Organization Websitehttp://www.lne.fr
Organization Country ISO CodeFR
Organization Country NameFrance
Analyte Category NameProteins