JCTLM Database: higher-order reference materials, methods and services
Details : C9RMS31 - HbA1c in Other, Whole blood
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C9RMS31 |
Reference Laboratory Status | National Metrology Institute |
Analyte Name | HbA1c |
IUPAC IFCC Number | QU60433 |
Applicable Matrices | lyophilized, fresh, or frozen human hemolysate |
Traceable to SI | Yes |
Principle of Analytical Procedure | ESI-LC/MS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | IFCC method for HbA1c (HPLC / MS-ESI) |
JCTLM Database Identifier Number | C1RMP_P16_MS |
Procedure Difference in Your Laboratory | YES |
Quantity | Amount-of-substance fraction of [HbA1c/(HbA1c+HbA0)] expressed in mmol/mol |
From (0) | 1 |
To (0) | 15 |
Unit (0) | % |
From (1) | 2 |
To (1) | 4 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is relative |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | Accredited |
Survey Organisation | IFCC RELA 2010 (Lab 105) |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | LNE's acceptable. |
Organization Acronym | LNE |
Organization Name | Laboratoire National de Métrologie et d'Essais |
Organization Website | http://www.lne.fr |
Organization Country ISO Code | FR |
Organization Country Name | France |
Analyte Category Name | Proteins |