JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_01 - Analysis of Ca by ICP-OES
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | CYC_II_L_I_RMP_01 |
| Analyte Name | calcium |
| IUPAC IFCC Number | CAS7440-70-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | freeze-dried or frozen serum |
| Identifier Number | Inorg-012 |
| Name | Analysis of Ca by ICP-OES |
| Organization that developed Validated Method | LGC |
| Measurement Techniques Used | ICP-OES |
| Comments 1 | Gravimetric preparation of primary standard using neat chemical |
| Quantity | Mass fraction |
| From (0) | 50 |
| To (0) | 140 |
| Unit (0) | µg/g |
| From (1) | 0.5 |
| To (1) | 2 |
| Unit (1) | % |
| Peer Reviewed Publication | Article accepted for publication by Talanta |
| Review Team Name | electrolytes |
| Analyte Category Name | Electrolytes |
| CCQM Key Comparison Report | Metrologia, 2004, 41, Tech. Suppl., 08003
|
| CCQM Key Comparison Report Hyperlink | http://stacks.iop.org/met/41/08003 |
| Measurement principle/technique | Inductively coupled plasma optical emission spectrometry |