JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_03 - CDC method for free glycerol in serum
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | CYC_II_L_I_RMP_03 |
Analyte Name | free glycerol |
IUPAC IFCC Number | free glycerol |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | fresh, frozen, lyophilized serum |
Name | CDC method for free glycerol in serum |
Organization that developed Validated Method | Centers for Disease Control, Atlanta, GA, USA |
Measurement Techniques Used | ID/GC/MS |
Comments 1 | trimethylsilyl derative |
Comments 2 | Calibrate with ACS reagent grade glycerol |
Quantity | Amount-of-substance concentration |
From (0) | 0.04 |
To (0) | 0.6 |
Unit (0) | mmol/l |
From (1) | 4 |
To (1) | 2 |
Unit (1) | % |
Peer Reviewed Publication | J. Chromatogr., 1992, 578, 1-7
|
How to Obtain Method Copy | Contact Parvin Waymack, pwaymack@cdc.gov for updated version |
Additional Comments | Used to measure free glycerol in CDC serum reference materials and evaluate glycerol-corrected vs non-corrected enzymatic results. Clin Chem 49, A165 (2003). % reative expanded unceratinty was calculated from QC data and value -assignment data from the last 7 years |
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Measurement principle/technique | Isotope dilution mass spectrometry |