JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_07 - LGC reference method for creatinine in serum by ID/LC/MS
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | CYC_II_L_I_RMP_07 |
| Analyte Name | creatinine |
| IUPAC IFCC Number | CAS60-27-5 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized or frozen serum |
| Name | LGC reference method for creatinine in serum by ID/LC/MS |
| Organization that developed Validated Method | LGC |
| Measurement Techniques Used | Isotope dilution LCMS |
| Comments 1 | Gravimetric preparation of primary standard using SRM 914a |
| Quantity | Mass fraction |
| From (0) | 5 |
| To (0) | 20 |
| Unit (0) | µg/g |
| From (1) | 0.3 |
| To (1) | 0.5 |
| Unit (1) | % |
| Peer Reviewed Publication | J.Chromatogr. B, 2003, 794, 125-136
|
| Review Team Name | metabolites and substrates |
| Analyte Category Name | Metabolites and substrates |
| CCQM Key Comparison Report | Metrologia, 2003, 40, Tech. Suppl., 08003
|
| CCQM Key Comparison Report Hyperlink | http://stacks.iop.org/met/40/08003 |
| Non-CCQM Key Comparison Report | IMEP-17 |
| Non-CCQM Key Comparison Report Hyperlink | |
|---|