JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_09 - NIST reference method for mercury in blood
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | CYC_II_L_I_RMP_09 |
Analyte Name | mercury |
IUPAC IFCC Number | CAS7439-97-6 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | frozen (liquid, not lyophilized) whole blood |
Identifier Number | 1 |
Name | NIST reference method for mercury in blood |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID/CV/ICPMS |
Quantity | Mass concentration |
From (0) | 0.1 |
To (0) | 30 |
Unit (0) | ng/g |
From (1) | 0.6 |
To (1) | 5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Anal. Chem., 2001, 73, 2190-2199
|
Review Team Name | non-electrolyte metals |
Analyte Category Name | Non-electrolyte metals |
Measurement principle/technique | Isotope dilution mass spectrometry |