Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : CYC_II_L_I_RMP_11 - NIST reference method for lead in blood

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberCYC_II_L_I_RMP_11
Analyte Namelead
IUPAC IFCC NumberCAS7439-92-1
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfrozen (liquid, not lyophilized) whole blood
Identifier Number3
NameNIST reference method for lead in blood
Organization that developed Validated MethodNIST
Measurement Techniques UsedID/ICPMS
QuantityMass concentration
From (1)0.24
To (1)5
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationFresenius J. Anal. Chem., 1995, 352, 203-208
Review Team Namenon-electrolyte metals
Analyte Category NameNon-electrolyte metals
CCQM Key Comparison ReportMetrologia, 2002, 39, Tech. Suppl., 08004
CCQM Key Comparison Report Hyperlinkhttp://www.iop.org/EJ/abstract/0026-1394/39/1A/22/
Measurement principle/techniqueIsotope dilution mass spectrometry