Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : CYC_II_L_I_RMP_12 - NIST reference method for lead in urine

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberCYC_II_L_I_RMP_12
Analyte Namelead
IUPAC IFCC NumberCAS7439-92-1
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen urine
Identifier Number4
NameNIST reference method for lead in urine
Organization that developed Validated MethodNIST
Measurement Techniques UsedID/ICPMS
QuantityMass concentration
From (1)0.24
To (1)5
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationFresenius J. Anal. Chem., 1995, 352, 203-208
Review Team Namenon-electrolyte metals
Analyte Category NameNon-electrolyte metals
CCQM Key Comparison ReportMetrologia, 2002, 39, Tech. Suppl., 08004
CCQM Key Comparison Report Hyperlinkhttp://www.iop.org/EJ/abstract/0026-1394/39/1A/22/
Measurement principle/techniqueIsotope dilution mass spectrometry