JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_13 - NIST reference method for cadmium in blood
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | CYC_II_L_I_RMP_13 |
| Analyte Name | cadmium |
| IUPAC IFCC Number | CAS7440-43-9 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | frozen (liquid, not lyophilized) whole blood |
| Identifier Number | 5 |
| Name | NIST reference method for cadmium in blood |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID/ICPMS |
| Quantity | Mass concentration |
| From (1) | 3 |
| To (1) | 5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Fresenius J. Anal. Chem., 2000, 368, 362-370
|
| Additional Comments | Used for certification of SRM 966, Frozen Whole Blood |
| Review Team Name | non-electrolyte metals |
| Analyte Category Name | Non-electrolyte metals |
| CCQM Key Comparison Report | Metrologia, 2002, 39, Tech. Suppl., 08004
|
| CCQM Key Comparison Report Hyperlink | http://www.iop.org/EJ/abstract/0026-1394/39/1A/22/ |
| Measurement principle/technique | Isotope dilution mass spectrometry |