JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_13 - NIST reference method for cadmium in blood
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | CYC_II_L_I_RMP_13 |
Analyte Name | cadmium |
IUPAC IFCC Number | CAS7440-43-9 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | frozen (liquid, not lyophilized) whole blood |
Identifier Number | 5 |
Name | NIST reference method for cadmium in blood |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID/ICPMS |
Quantity | Mass concentration |
From (1) | 3 |
To (1) | 5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Fresenius J. Anal. Chem., 2000, 368, 362-370
|
Additional Comments | Used for certification of SRM 966, Frozen Whole Blood |
Review Team Name | non-electrolyte metals |
Analyte Category Name | Non-electrolyte metals |
CCQM Key Comparison Report | Metrologia, 2002, 39, Tech. Suppl., 08004
|
CCQM Key Comparison Report Hyperlink | http://www.iop.org/EJ/abstract/0026-1394/39/1A/22/ |
Measurement principle/technique | Isotope dilution mass spectrometry |