Details : CYC_II_L_I_RMP_13 - NIST reference method for cadmium in blood

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberCYC_II_L_I_RMP_13
Analyte Namecadmium
IUPAC IFCC NumberCAS7440-43-9
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfrozen (liquid, not lyophilized) whole blood
Identifier Number5
NameNIST reference method for cadmium in blood
Organization that developed Validated MethodNIST
Measurement Techniques UsedID/ICPMS
QuantityMass concentration
From (1)3
To (1)5
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationFresenius J. Anal. Chem., 2000, 368, 362-370
Additional CommentsUsed for certification of SRM 966, Frozen Whole Blood
Review Team Namenon-electrolyte metals
Analyte Category NameNon-electrolyte metals
CCQM Key Comparison ReportMetrologia, 2002, 39, Tech. Suppl., 08004
CCQM Key Comparison Report Hyperlinkhttp://www.iop.org/EJ/abstract/0026-1394/39/1A/22/