JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_22 - NIST instrumental neutron activation analysis method for cobalt in urine
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | CYC_II_L_I_RMP_22 |
| Analyte Name | cobalt |
| IUPAC IFCC Number | CAS7440-48-4 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | frozen (liquid) or lyophilized urine |
| Identifier Number | 28 |
| Name | NIST instrumental neutron activation analysis method for cobalt in urine |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | INAA |
| Comments 1 | Validate with NIST SRM 1577a and 1577b Bovine Liver |
| Quantity | Mass concentration |
| From (1) | 3 |
| To (1) | 15 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | J. Radioanal. Nucl. Chem., 1998, 124, 47-63
Biol. Tr. Elem. Res., 1984, 6, 31-49
|
| Review Team Name | non-electrolyte metals |
| Analyte Category Name | Non-electrolyte metals |
| Measurement principle/technique | Instrumental neutron activation analysis |