JCTLM Database: higher-order reference materials, methods and services
Details : CYC_II_L_I_RMP_24 - NIST LC/MS/MS reference method for total triiodothyronine
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | CYC_II_L_I_RMP_24 |
Analyte Name | total triiodothyronine |
IUPAC IFCC Number | CAS6893-02-3 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | human serum |
Name | NIST LC/MS/MS reference method for total triiodothyronine |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID-LC/MS/MS |
Comments 1 | Corrected for conversion of T4 to T3 in sample prep |
Quantity | Mass concentration |
From (0) | 1 |
To (0) | 3 |
Unit (0) | ng/ml |
From (1) | 3 |
To (1) | 5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Anal. Chem. 2004, 76, 5092-5096
|
How to Obtain Method Copy | susan.tai@nist.gov |
Review Team Name | non-peptide hormones |
Analyte Category Name | Non-peptide hormones |
Non-CCQM Key Comparison Report | See reference cited for comparability assessment study |
Non-CCQM Key Comparison Report Hyperlink | |
---|