JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 1 - University of Ghent reference measurement procedure for creatinine
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 1 |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | University of Ghent reference measurement procedure for creatinine |
Organization that developed Validated Method | University of Ghent |
Measurement Techniques Used | ID/GC/MS |
Comments 1 | tert-butyl-dimethylsilyl derivative |
Comments 2 | Calibrated with SRM914a, high purity creatinine; accuracy control with SRM 909b, CRM 573-575 and DGKC-sera [note: DGKC = Deutsche Gesellschaft fuer Klinische Chemie] |
Quantity | Amount-of-substance concentration |
From (0) | 15 |
To (0) | 500 |
Unit (0) | µmol/l |
From (1) | 1 |
To (1) | 2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Clin. Chem., 1993, 39,1001-1006
Clin. Chem., 1993, 39, 993-1000
Eur. J. Clin. Chem. Clin. Biochem., 1996, 34, 853-860
Clin. Chem., 1995, 41 , 995-1003
|
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Non-CCQM Key Comparison Report | EUROMET 563 |
Non-CCQM Key Comparison Report Hyperlink | |
---|