Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : NRMeth 1 - University of Ghent reference measurement procedure for creatinine

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberNRMeth 1
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameUniversity of Ghent reference measurement procedure for creatinine
Organization that developed Validated MethodUniversity of Ghent
Measurement Techniques UsedID/GC/MS
Comments 1tert-butyl-dimethylsilyl derivative
Comments 2Calibrated with SRM914a, high purity creatinine; accuracy control with SRM 909b, CRM 573-575 and DGKC-sera [note: DGKC = Deutsche Gesellschaft fuer Klinische Chemie]
QuantityAmount-of-substance concentration
From (0)15
To (0)500
Unit (0)µmol/l
From (1)1
To (1)2
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationClin. Chem., 1993, 39,1001-1006 Clin. Chem., 1993, 39, 993-1000 Eur. J. Clin. Chem. Clin. Biochem., 1996, 34, 853-860 Clin. Chem., 1995, 41 , 995-1003
Review Team Namemetabolites and substrates
Analyte Category NameMetabolites and substrates
Non-CCQM Key Comparison ReportEUROMET 563
Non-CCQM Key Comparison Report Hyperlink
Measurement principle/techniqueIsotope dilution mass spectrometry