Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : NRMeth 14 - University of Ghent ion chromatographic reference method for serum sodium

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberNRMeth 14
Analyte Namesodium
IUPAC IFCC NumberCAS7440-23-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameUniversity of Ghent ion chromatographic reference method for serum sodium
Organization that developed Validated MethodUniversity of Ghent
Measurement Techniques UsedIon chromatography
Comments 2Calibrated with SRM 919a, high purity sodium chloride; accuracy control with SRM 909b and DGKC-sera
QuantityAmount-of-substance concentration
From (0)110
To (0)170
Unit (0)mmol/l
From (1)1
To (1)2
Unit (1)%
Peer Reviewed PublicationJ. Chromatogr. A, 1995, 706, 443-450 Clin. Biochem., 1996, 29, 501-508 J. Chromatogr. A, 1997, 789, 557-568
Comments to be published via JCTLM DBThe method measures the amount of sodium per volume of serum. This provides results suitable for traceability of indirect ion-sensitive electrode measurements
Review Team Nameelectrolytes
Analyte Category NameElectrolytes
Measurement principle/techniqueIon chromatography