JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 16 - University of Ghent ion chromatographic reference method for serum calcium
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 16 |
Analyte Name | calcium |
IUPAC IFCC Number | CAS7440-70-2 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | University of Ghent ion chromatographic reference method for serum calcium |
Organization that developed Validated Method | University of Ghent |
Measurement Techniques Used | Ion chromatography |
Comments 2 | Calibrated with SRM 915a, high purity calcium carbonate; accuracy control with SRM 909b, CRM 303 and 304 and DGKC-sera |
Quantity | Amount-of-substance concentration |
From (0) | 1.25 |
To (0) | 3 |
Unit (0) | mmol/l |
From (1) | 1 |
To (1) | 2 |
Unit (1) | % |
Peer Reviewed Publication | Anal. Chem., 1994, 66, 2404-2408
Clin. Biochem., 1996, 29, 501-508
J. Chromatogr. A,1997, 789, 557-568
Eur. J. Clin. Chem. Clin. Biochem., 1997, 35, 297-300
Scand. J. Clin. Lab. Invest., 1998, 58, 229-240
|
Review Team Name | electrolytes |
Analyte Category Name | Electrolytes |
Measurement principle/technique | Ion chromatography |