Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : NRMeth 16 - University of Ghent ion chromatographic reference method for serum calcium

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberNRMeth 16
Analyte Namecalcium
IUPAC IFCC NumberCAS7440-70-2
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameUniversity of Ghent ion chromatographic reference method for serum calcium
Organization that developed Validated MethodUniversity of Ghent
Measurement Techniques UsedIon chromatography
Comments 2Calibrated with SRM 915a, high purity calcium carbonate; accuracy control with SRM 909b, CRM 303 and 304 and DGKC-sera
QuantityAmount-of-substance concentration
From (0)1.25
To (0)3
Unit (0)mmol/l
From (1)1
To (1)2
Unit (1)%
Peer Reviewed PublicationAnal. Chem., 1994, 66, 2404-2408 Clin. Biochem., 1996, 29, 501-508 J. Chromatogr. A,1997, 789, 557-568 Eur. J. Clin. Chem. Clin. Biochem., 1997, 35, 297-300 Scand. J. Clin. Lab. Invest., 1998, 58, 229-240
Review Team Nameelectrolytes
Analyte Category NameElectrolytes
Measurement principle/techniqueIon chromatography