JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 19 - NIST ID-TIMS method for serum calcium
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 19 |
| Analyte Name | calcium |
| IUPAC IFCC Number | CAS7440-70-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | human serum |
| Name | NIST ID-TIMS method for serum calcium |
| Organization that developed Validated Method | NCCLS/NRSCL |
| Measurement Techniques Used | IDMS |
| Comments 1 | see Anal Chem 44:2291 |
| Comments 2 | See NBS special Publication 260-36 |
| Quantity | Amount-of-substance concentration |
| From (0) | 1.8 |
| To (0) | 2.9 |
| Unit (0) | mmol/l |
| From (1) | 0.0037 |
| To (1) | 0.006 |
| Unit (1) | mmol/l |
| Peer Reviewed Publication | NBS special publication 260-36
|
| How to Obtain Method Copy | Contact NCCLS |
| Review Team Name | electrolytes |
| Analyte Category Name | Electrolytes |
| Non-CCQM Key Comparison Report | See reference cited for comparability assessment study |
| Non-CCQM Key Comparison Report Hyperlink | |
|---|