JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 4 - University of Ghent reference method for glucose
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 4 |
Analyte Name | glucose |
IUPAC IFCC Number | CAS50-99-7 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | University of Ghent reference method for glucose |
Organization that developed Validated Method | University of Ghent |
Measurement Techniques Used | ID/GC/MS |
Comments 1 | Nitrilopentaacetate derivative |
Comments 2 | Calibrated with SRM 917b, high purity glucose; accuracy control with DGKC-sera |
Quantity | Amount-of-substance concentration |
From (0) | 1 |
To (0) | 20 |
Unit (0) | mmol/l |
From (1) | 1 |
To (1) | 2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Clin. Chem., 1993, 39, 1001-1006
Clin. Chem., 1993, 39, 993-1000
Eur. J. Clin. Chem. Clin. Biochem., 1996, 34, 853-860
|
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Non-CCQM Key Comparison Report | EUROMET 563 |
Non-CCQM Key Comparison Report Hyperlink | |
---|