Details : NRMeth 50 - DGKC reference method for uric acid in serum or plasma
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 50 |
Analyte Name | uric acid |
IUPAC IFCC Number | CAS69-93-2 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum or plasma |
Name | DGKC reference method for uric acid in serum or plasma |
Organization that developed Validated Method | German Society of Clinical Chemistry (DGKC) |
Measurement Techniques Used | ID/GC/MS |
Comments 1 | Tert-butyl-di-methyl-silyl derivative |
Comments 2 | Calibrate with SRM 913, high purity uric acid |
Quantity | Amount-of-substance concentration |
From (0) | 100 |
To (0) | 700 |
Unit (0) | µmol/l |
From (1) | 0.5 |
To (1) | 1.5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Siekmann et al., J. Clin. Chem. Clin. Biochem., 1985, 23, 129-135
|
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Non-CCQM Key Comparison Report | Benchmarked with PTB reference method for uric acid in serum or plasma |
Non-CCQM Key Comparison Report Hyperlink | |
Non-CCQM Key Comparison Report | for information, contact lothar.siekmann@ukb.uni-bonn.de |
Non-CCQM Key Comparison Report Hyperlink | |
Measurement principle/technique | Isotope dilution mass spectrometry |