JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 58 - DGKC reference method for serum progesterone
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 58 |
| Analyte Name | progesterone |
| IUPAC IFCC Number | CAS57-83-0 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum or plasma |
| Name | DGKC reference method for serum progesterone |
| Organization that developed Validated Method | German Society of Clinical Chemistry (DGKC) |
| Measurement Techniques Used | ID/GC/MS |
| Comments 1 | Heptafluorobutyric ester |
| Comments 2 | Calibrate with Progesterone (internally certified) |
| Quantity | Amount-of-substance concentration |
| From (0) | 1.5 |
| To (0) | 90 |
| Unit (0) | nmol/l |
| From (1) | 1 |
| To (1) | 3 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Siekmann et al., J. Steroid Biochem., 1979, 11, 117-123
|
| How to Obtain Method Copy | e-mail: lothar.siekmann@ukb.uni-bonn.de |
| Review Team Name | non-peptide hormones |
| Analyte Category Name | Non-peptide hormones |
| Non-CCQM Key Comparison Report | The certification of progesterone in two lyophilized serum materials, BCR CRM 347 and 348: Report EUR 12282 EN, 1989 |
| Non-CCQM Key Comparison Report Hyperlink | |
|---|