Details : NRMeth 66 - IFCC reference measurement procedure (37 °C) for LDH
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 66 |
| Analyte Name | lactate dehydrogenase (LDH) |
| IUPAC IFCC Number | EC 1.1.1.27 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | Defined procedure |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | IFCC reference measurement procedure (37 °C) for LDH |
| Organization that developed Validated Method | IFCC |
| Measurement Techniques Used | Kinetic spectrophotometry |
| Comments 1 | Reference Measurement Procedure approved by IFCC Mail Ballot |
| Quantity | Catalytic amount |
| From (0) | 0.068 |
| To (0) | 10 |
| Unit (0) | µkat/l |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | 2.2 % to open |
| Peer Reviewed Publication | Clin. Chem. Lab. Med., 2002, 40, 643-648
|
| Measurand Clearly Defined | By procedure itself |
| Multiple Testing Sites Not Required to obtain a Single Value | No |
| Review Team Name | enzymes |
| Analyte Category Name | Enzymes |
| Non-CCQM Key Comparison Report | Clin. Chem. Lab. Med., 2002, 40, 739-745
|
| Measurement principle/technique | Spectrophotometry |