JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 73 - NIST ID-ICP/MS method serum potassium
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 73 |
Analyte Name | potassium |
IUPAC IFCC Number | CAS7440-09-7 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | NIST ID-ICP/MS method serum potassium |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID/ICP-MS |
Comments 1 | K-41, cold plasma |
Comments 2 | Calibrate with SRM 918a |
Quantity | Amount-of-substance concentration |
From (0) | 1 |
To (0) | 8 |
Unit (0) | mmol/l |
From (1) | 0.4 |
To (1) | 0.4 |
Unit (1) | % |
Peer Reviewed Publication | J. Anal. At. Spectrom., 2002, 17, 469-477
|
Review Team Name | electrolytes |
Analyte Category Name | Electrolytes |
Measurement principle/technique | Isotope dilution mass spectrometry |