JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 74 - NIST ID-TIMS method for serum lithium
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 74 |
| Analyte Name | lithium |
| IUPAC IFCC Number | CAS7439-93-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | NIST ID-TIMS method for serum lithium |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID/TIMS |
| Comments 1 | Li-6, + triple filament as Li phosphate |
| Comments 2 | Calibrate with SRM 924a |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.1 |
| To (0) | 5 |
| Unit (0) | mmol/l |
| From (1) | 0.3 |
| To (1) | 0.4 |
| Unit (1) | % |
| Peer Reviewed Publication | Fres. J. Anal. Chem., 1998, 361, 71-80
|
| Review Team Name | electrolytes |
| Analyte Category Name | Electrolytes |
| Measurement principle/technique | Isotope dilution mass spectrometry |