JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 75 - NIST ID/ICP-MS method for serum calcium
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 75 |
Analyte Name | calcium |
IUPAC IFCC Number | CAS7440-70-2 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | NIST ID/ICP-MS method for serum calcium |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID/ICP-MS |
Comments 1 | Ca-42, cold plasma |
Comments 2 | Calibrate with SRM 915a |
Quantity | Mass concentration |
From (0) | 5 |
To (0) | 15 |
Unit (0) | mg/dl |
From (1) | 0.4 |
To (1) | 0.4 |
Unit (1) | % |
Peer Reviewed Publication | J. Anal. At. Spectrom., 2002, 17, 469-477
|
Review Team Name | electrolytes |
Analyte Category Name | Electrolytes |
CCQM Key Comparison Report | Metrologia, 2004, 41, Tech. Suppl., 08003
|
CCQM Key Comparison Report Hyperlink | http://stacks.iop.org/met/41/08003 |
Measurement principle/technique | Isotope dilution mass spectrometry |