JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 75 - NIST ID/ICP-MS method for serum calcium
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 75 |
| Analyte Name | calcium |
| IUPAC IFCC Number | CAS7440-70-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | NIST ID/ICP-MS method for serum calcium |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID/ICP-MS |
| Comments 1 | Ca-42, cold plasma |
| Comments 2 | Calibrate with SRM 915a |
| Quantity | Mass concentration |
| From (0) | 5 |
| To (0) | 15 |
| Unit (0) | mg/dl |
| From (1) | 0.4 |
| To (1) | 0.4 |
| Unit (1) | % |
| Peer Reviewed Publication | J. Anal. At. Spectrom., 2002, 17, 469-477
|
| Review Team Name | electrolytes |
| Analyte Category Name | Electrolytes |
| CCQM Key Comparison Report | Metrologia, 2004, 41, Tech. Suppl., 08003
|
| CCQM Key Comparison Report Hyperlink | http://stacks.iop.org/met/41/08003 |
| Measurement principle/technique | Isotope dilution mass spectrometry |