Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : NRMeth 78 - NIST coulometric method for serum chloride

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberNRMeth 78
Analyte Namechloride
IUPAC IFCC NumberCAS16887-00-6
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameNIST coulometric method for serum chloride
Organization that developed Validated MethodNIST
Measurement Techniques UsedCoulometry
Comments 1Titration of electro-generated Ag+
Comments 2Cl Calculated from Faraday's Laws
QuantityAmount-of-substance concentration
From (0)75
To (0)150
Unit (0)mmol/l
From (1)0.1
To (1)0.2
Unit (1)%
Peer Reviewed PublicationFres. J. Anal. Chem., 1998, 361, 71-80
Review Team Nameelectrolytes
Analyte Category NameElectrolytes
Measurement principle/techniqueCoulometry