JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 84 - NIST reference method for serum uric acid
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | NRMeth 84 |
| Analyte Name | uric acid |
| IUPAC IFCC Number | CAS69-93-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | NIST reference method for serum uric acid |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID/GC/MS |
| Comments 1 | tetrakis-(tert-butyldimethylsilyl) derivative |
| Comments 2 | Calibrate with SRM 913a Uric Acid |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.2 |
| To (0) | 0.8 |
| Unit (0) | mmol/l |
| From (1) | 1 |
| To (1) | 2.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Anal. Chem., 1990, 62, 2173-2177
|
| Review Team Name | metabolites and substrates |
| Analyte Category Name | Metabolites and substrates |
| Measurement principle/technique | Isotope dilution mass spectrometry |