JCTLM Database: higher-order reference materials, methods and services
Details : NRMeth 90 - NIST LC/MS reference method for serum folate (5-methyltetrahydrofolic acid)
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | NRMeth 90 |
Analyte Name | 5-methyltetrahydrofolic acid |
IUPAC IFCC Number | CAS68792-52-9 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | NIST LC/MS reference method for serum folate (5-methyltetrahydrofolic acid) |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID/LC/MS |
Comments 1 | no derivative |
Comments 2 | Calibrate with commercial material characterized at NIST |
Quantity | Mass concentration |
From (0) | 7 |
To (0) | 19 |
Unit (0) | ng/ml |
From (1) | 3 |
To (1) | 9 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Anal. Biochem., 2003, 313, 117-127
|
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Measurement principle/technique | Isotope dilution mass spectrometry |