Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : NRMeth40_Estrone - Mayo reference measurement method for estrone in blood plasma

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberNRMeth40_Estrone
Analyte Nameestrone
IUPAC IFCC NumberCAS50-28-2
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceshuman plasma
NameMayo reference measurement method for estrone in blood plasma
Organization that developed Validated MethodMayo Foundation
Measurement Techniques UsedLC-MS/MS
Comments 1Ion Pair 271/145 normalized to d5-estradiol ion pair 276/187
Comments 2C18 column (Supleco Supelcosil)
QuantityAmount-of-substance concentration
From (0)0.037
To (0)7.4
Unit (0)nmol/l
Peer Reviewed PublicationLiquid Chromatography Tandem Mass Spectrometry Assay for Simultaneous Measurement of Estradiol and Estrone in Human Plasma, Nelson R et al., Clin. Chem., 2004, 50, 373
Review Team Namenon-peptide hormones
Analyte Category NameNon-peptide hormones
Measurement principle/techniqueLiquid chromatography mass spectrometry